Selected System Specifications Agilent 1260 Infinity Isocratic lc system Pump principle




Yüklə 14.94 Kb.
tarix11.03.2016
ölçüsü14.94 Kb.
Selected System Specifications - Agilent 1260 Infinity Isocratic LC System

Pump principle

Single channel



Flow rate range

0.05 – 5 mL/min at 600 bar


5 –10 mL/min at 200 bar

Maximum pressure

600 bar


Injection principle

Flow-through design, no overfill (drawn volume equals injected volume). Autosamplers are available for vials only or for vials and micro-titer plates. Manual injector up to 600 bar available.



Injection volumes

0.1 – 100 µL, variable in 0.1 µL steps



Column dimensions

Length: up to 300 mm, ID: 0.05 – 8 mm



Temperature range

10 degrees below ambient to 80 °C (or to 100 °C with 1290 Infinity Thermostatted Column Compartment)



Temperature stability

< ± 0.15 °C (< ± 0.05 °C with 1290 Infinity Thermostatted Column Compartment)

Detection data rate

80 Hz for UV, 74 Hz for fluorescence, 60 Hz for evaporative light scattering detection, higher data rates available with 1290 Infinity detectors



UV noise level

As low as 0.6 uAU/cm (1260 Infinity DAD, 60 mm cell)



Maintenance and system test

Front access to all maintainable parts.


Maintenance instructions through multimedia CD.
Time for full system test (OQ/PV) < 4 hours.

Standard GLP features

Early Maintenance Feedback (EMF) tracks lamp burn time, usage, number of injections with limits and feedback messages



Extended GLP features
(for certain systems, refer to specifications of individual modules)

Data Recovery Card (DRC) for data security, preventing data loss in case of communication breakdowns.


Radio Frequency Identification (RFID) for data traceability. RFID tags for flow cells and UV lamp allow unambiguous traceability of measurement conditions.

System control

Through local workstation software, client/server network data system and/or local handheld control module (1200 Infinity Series Instant Pilot). Default control: LAN / CAN.



Agilent 1260 Infinity Quaternary LC System - Selected System Specifications

Pump principle

Low pressure quaternary mixing, integrated high performance degassing unit



Flow rate range

0.05 - 5 mL/min at 600 bar


5 - 10 mL/min at 200 bar

Maximum pressure

600 bar


Injection principle

Flow through design, no overfill (drawn = injected volume). Vial-only autosampler and vial and micro-titer plate autosampler available.



Injection volumes

0.1 - 100 µL, variable in 0.1 µL steps, higher injection volumes optionally available



Column dimensions

Length: up to 300 mm, ID: 0.05 - 8 mm



Temperature range

10 °C below ambient - 80 °C


With optional 1290 Infinity Thermostatted Column Compartment up to 100 °C

Temperature stability

< ± 0.15 °C
With optional 1290 Infinity Thermostatted Column Compartment
< ± 0.05 °C

Detection data rate

80 Hz for UV, 74 Hz for fluorescence, 60 Hz for evaporative light scattering detection, higher data rates available with 1290 Infinity detectors



UV noise level

As low as 0.6 uAU/cm (1260 Infinity DAD, 60 mm cell)



Maintenance and system test

Front access to all maintenance parts


Maintenance instructions through multimedia CD-ROM
Time for full system test (OQ/PV) < 4 hours

Standard GLP features

Early maintenance feedback-EMF


(tracks lamp burn time, usage, number of injections with limits and feedback messages)

Extended GLP features
(for certain systems, refer to specifications of individual modules)

Data Recovery Card (DRC) for data security (prevents data losses in case of communication breakdowns – depends on actually chosen detector)


Radio Frequency Identification (RFID) for data traceability. (RFID tags for flow cells and UV lamp allow unambiguous traceability of measurement conditions)

System control

Through local workstation software, client/server network data system and/or local handheld control module (1200 Infinity Series Instant Pilot). Default control: LAN / CAN.


Verilənlər bazası müəlliflik hüququ ilə müdafiə olunur ©azrefs.org 2016