Rename and Extension of Cardiac Trigger Macro Status: Letter Ballot

DICOM Correction Item




Correction Number CP-593

Log Summary: Rename and Extension of Cardiac Trigger Macro

Type of Modification

Clarification / Addition / Correction



Name of Standard

PS 3.3 - 2004 + Supplement 83



Rationale for Correction:

This change proposal proposes registration of respiratory triggers for respiratory and cardiac/respiratory examination for the Enhanced MR, Spectroscopy, CT multi-frame IOD and Enhanced XA/XRF.

To fulfill the requirements the next extensions and changes are proposed:

1. Extension of the Cardiac Trigger Functional Group.

2. Introduction of a new Functional group for Respiratory Triggers.


Sections of documents affected:

PS 3.3 Sections A.36.2.4, A.36.3.4, A.38.1.4, C.7.6.16.2, C.7.6.18.2, C.8.15.2.1.1.3 and C.8.16.1.3

PS 3.3, Part 6.

Supplement 83 Sections A.X.4 and A.Y.4



Correction Wording:

Item 1: Add new functional group to PS 3.3, Section A.36.2.4

A.36.2.4 Enhanced MR Image Functional Group Macros


Table A.36-2 specifies the use of the Functional Group macros used in the Multi-frame Functional Groups Module for the Enhanced MR Image IOD.

Table A.36-2


ENHANCED MR IMAGE FUNCTIONAL GROUP MACROS

Functional Group Macro

Section

Usage









Cardiac Trigger

C.7.6.16.2.7

C - Required if Cardiac Synchronization Technique (0018,9037) equals other than NONE and if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise.









Contrast/Bolus Usage

C.7.6.16.2.12

C – Required if Contrast/Bolus Agent Sequence (0018,0012) is used. May not be used as a Shared Functional Group

Respiratory Trigger

C.7.6.16.2.X

C – Required if Respiratory Motion Compensation Technique (0018,9170) equals other than NONE, REALTIME or BREATH_HOLD and if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise.









Item 2: Add new functional group to PS 3.3, Section A.36.3.4

A.36.3.4 MR Spectroscopy Functional Group Macros


Table A.36-4 specifies the use of the Functional Group macros used in the Multi-frame Functional Groups Module for the MR Spectroscopy IOD.

Table A.36-4


MR SPECTROSCOPY FUNCTIONAL GROUP MACROS

Functional Group Macro

Section

Usage









Cardiac Trigger

C.7.6.16.2.7

C - Required if Cardiac Synchronization Technique (0018,9037) equals other than NONE and if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise.









Contrast/Bolus Usage

C.7.6.16.2.12

C – Required if Contrast/Bolus Agent Sequence (0018,0012) is used. May not be used as a Shared Functional Group

Respiratory Trigger

C.7.6.16.2.X

C – Required if Respiratory Motion Compensation Technique (0018,9170) equals other than NONE, REALTIME or BREATH_HOLD and if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise.









Item 3: Add new functional group to PS 3.3, Section A.38.1.4

A.38.1.4 Enhanced CT Image Functional Group Macros


Table A.38-2 specifies the use of the Functional Group macros used in the Multi-frame Functional Group Module for the Enhanced CT Image IOD.

Table A.38-2


ENHANCED CT IMAGE FUNCTIONAL GROUP MACROS

Functional Group Macro

Section

Usage
















Cardiac Trigger

C.7.6.16.2.7

C - Required if Cardiac Synchronization Technique (0018,9037) equals other than NONE and if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise.







Contrast/Bolus Usage

C.7.6.16.2.12

C – Required if Contrast/Bolus Agent Sequence (0018,0012) is used. May not be used as a Shared Functional Group

Respiratory Trigger

C.7.6.16.2.X

C – Required if Respiratory Motion Compensation Technique (0018,9170) equals other than NONE, REALTIME or BREATH_HOLD and if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise.







Item 4: Add new functional group to Supplement 83, Section A.X.4

A.X.4 Enhanced XA Image Functional Group Macros


Table A.X-2 specifies the use of the Functional Group macros used in the Multi-frame Functional Groups Module for the Enhanced XA Image IOD.

Table A.X-2


ENHANCED XA IMAGE FUNCTIONAL GROUP MACROS

Functional Group Macro

Section

Usage









Cardiac Trigger

C.7.6.16.2.7

U









Frame Display Shutter

C.7.6.16.2.X5

U

Respiratory Trigger

C.7.6.16.2.X

U









Item 5: Add new functional group to Supplement 83, Section A.Y.4

A.Y.4 Enhanced XRF Image Functional Group Macros


Table A.Y-2 specifies the use of the Functional Group macros used in the Multi-frame Functional Groups Module for the Enhanced XRF Image IOD.

Table A.Y-2


ENHANCED XRF IMAGE FUNCTIONAL GROUP MACROS

Functional Group Macro

Section

Usage









Cardiac Trigger

C.7.6.16.2.7

U









Frame Display Shutter

C.7.6.16.2.X5

U

Respiratory Trigger

C.7.6.16.2.X

U









Item 6: Add new attributes to PS 3., Section C.7.6.16.2.7 and modify Figure C.7.6.16-5
C.7.6.16.2.7 Cardiac Trigger Macro

Table C.7.6.16-8 specifies the attributes of the Cardiac Trigger Functional Group macro.

Table C.7.6.16-8


CARDIAC TRIGGER Macro ATTRIBUTES

Attribute Name

Tag

Type

Attribute Description

Cardiac Trigger Sequence

(0018,9118)

1

Identifies cardiac trigger delay for this frame. Only a single Item shall be permitted in this sequence.

>Cardiac Trigger Delay Time

(0020,9153)

1

Trigger delay time in ms from the previous R-peak to the value of the Frame Reference Datetime (0018,9151). See C.7.6.16.2.7.1 for further explanation.

>R – R Interval Time Measured

(0020,93x1)

1C

Measured R-peak – R-peak interval time in ms for the cardiac cycle in which this frame occurs. See C.7.6.16.2.7.1 for further explanation.

Required if Cardiac Synchronization Technique (0018,9037) equals other than NONE or REALTIME. May be present otherwise.

Item 7: Add new section to PS 3.3, Section C.7.6.16.2
C.7.6.16.2.X Respiratory Trigger Macro

Table C.7.6.16-X specifies the attributes of the Respiratory Trigger Functional Group macro.

Table C.7.6.16-X


RESPIRATORY TRIGGER Macro ATTRIBUTES

Attribute Name

Tag

Type

Attribute Description

Respiratory Trigger Sequence

(0018,93x3)

1

Identifies respiratory trigger delay for this frame. Only a single Item shall be permitted in this sequence.

>Respiratory Interval Time

(0020,93x4)

1

Measured interval time in ms from maximum respiration peak to the next peak for the respiratory cycle in which this frame occurs. See C.7.6.16.2.X.1 for further explanation.

Required if (Respiratory Motion Compensation Technique (0018,9170)) equals other than NONE or REALTIME) and (Respiratory Signal Source (0018,9171) is BELT). May be present otherwise.



>Respiratory Trigger Delay Time

(0020,93x5)

1

Trigger delay time in ms from the previous maximum value of respiration to the value of the Frame Reference Datetime (0018,9151). See C.7.6.16.2.X.1 for further explanation.


C.7.6.16.2.X.1 Relationship of Respiratory Timing Attributes

Figure C.7.6.16-X depicts the usage.

Figure C.7.6.16-X
Respiratory Timing Tags

Item 8: Add Respiration Trigger Delay Threshold to Table C.7.6.18-2


C.7.6.18.2 Respiratory Synchronization Module

Table C7.6.18-2 specifies the attributes of the Respiratory Synchronization Module.

Table C.7.6.18-2


RESPIRATORY SYNCHRONIZATION MODULE ATTRIBUTES

Attribute Name

Tag

Type

Attribute Description

Respiratory Signal Source

(0018,9171)

1C

Signal source from which respiratory motion is derived.

Defined Terms:

NONE

BELT


NASAL_PROBE

CO2_SENSOR

NAVIGATOR = MR navigator and organ edge detection

MR_PHASE = phase (of center k-space line)

ECG = baseline demodulation of the ECG

Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED and Respiratory Motion Compensation Technique (0018,9170) equals other than NONE.

Otherwise may be present if Image Type (0008,0008) Value 1 is DERIVED and Respiratory Motion Compensation Technique (0018,9170 equals other than NONE.


Respiratory Trigger Delay Threshold

(0020,93x6)

1C

Respiratory trigger threshold in percent of the chest expansion for the frame relative to the last Respiratory-Peak. See C.7.6.16.2.X.1 for further explanation.

Required if Respiratory Motion Compensation Technique (0018,9170) equals other than NONE, REALTIME or BREATH_HOLD and if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise.

Item 9: Move from PS 3.3, Table C.8-155 image and frame type Values 3 to Common Table and add new terms
C.8.15.2.1.1.3 Image Flavor

Table C.8-115 specifies the Defined Terms for CT additional to those defined in C.8.16.1.3 for Value 3 for Image Type (0008,0008) and Frame Type (0008,9007).

Table C.8-115


IMAGE TYPE AND FRAME TYPE VALUE 3 FOR CT

Defined Term Name

Defined Term Description

ATTENUATION

Collected for the purpose of performing attenuation corrections (e.g. PET attenuation correction)

CARDIAC

Cardiac images

CARDIAC_GATED

Cardiac gated images

REFERENCE

Collected for anatomical reference for PET or SPECT


C.8.16.1.3 Image Flavor

Table C.8-115


IMAGE TYPE AND FRAME TYPE VALUE 3 COMMON

Defined Term Name

Defined Term Description

ANGIO

Collected for the purpose of angiography

CARDIAC

Images of the heart

CARDIAC_GATED

Cardiac gated images, other than of the heart

CARDRESP_GATED

Cardiac and respiratory gated images






POST_CONTRAST

Collected during or after contrast was administered

RESP_GATED

Respiratory gated images












Item 10: Change and add new attributes to PS 3.6, Section 6

Tag

Name

VR

VM




(0008,9307)

Trigger Method

CS

1




(0020,9153)

Cardiac Trigger Delay Time

FD

1




(0020,93x1)

R – R Interval Time Measured

FD

1




(0020,93x3)

Respiratory Trigger Sequence

SQ

1




(0020,93x4)

Respiratory Interval Time

FD

1




(0020,93x5)

Respiratory Trigger Delay Time

FD

1




(0020,93x6)

Respiratory Trigger Delay Threshold

FD

1