Clarify Breast Implant Present description Status: Letter Ballot




Yüklə 18.37 Kb.
tarix14.04.2016
ölçüsü18.37 Kb.

CP-989 Date: 2010/01/18

Clarify Breast Implant Present description Status: Letter Ballot



DICOM Correction Item

Correction Number CP-989

Log Summary: Clarify Breast Implant Present description

Type of Modification

Clarification



Name of Standard

PS 3.3 2009



Rationale for Correction:

It is not clear to implementers when the value of Breast Implant Present (0028,1300) should be “YES” versus “NO”. The current Attribute Description is vague.



Sections of documents affected

PS 3.3, C.8.11.7, C.8.21.6



Correction Wording:


Modify PS 3.3, Table C.8-74 as follows:

C.8.11.7 Mammography Image Module


Table C.8.74 contains IOD Attributes that describe a Digital Mammography X-Ray Image including its acquisition and positioning.

Table C.8-74


MAMMOGRAPHY IMAGE MODULE ATTRIBUTES

Attribute Name

Tag

Type

Attribute Description












Breast Implant Present

(0028,1300)

3

Whether or not the imaged breast contains a breast implant is present, regardless of the visibility of a breast implant in the Pixel Data. Enumerated Values:

YES
NO



Note: The value is expected to be YES for all images acquired on a breast that contains a breast implant, even when a breast implant is displaced during image acquisition.













Modify PS 3.3, Table C.8.21.6-1 as follows:

C.8.21.6 Breast View Module


Table C.8.21.6-1 contains IOD Attributes that describe the view of a Breast Tomosynthesis Image.

Table C.8.21.6-1


BREAST VIEW MODULE ATTRIBUTES

Attribute Name

Tag

Type

Attribute Description












Breast Implant Present

(0028,1300)

1C

Whether or not the imaged breast contains a breast implant is present, regardless of the visibility of a breast implant in the Pixel Data. Enumerated Values:

YES
NO



Required if Modality (0008,0060) is MG. May be present otherwise.

Note: The value is expected to be YES for all images acquired on a breast that contains a breast implant, even when a breast implant is displaced during image acquisition.